8–10 mins Rapid Readout Time | 0.5 µg/kg Strict Trace LOD (0.5 ppb) | 10 Tests Complete Kit Pack | 18 Months Shelf Life (4–30°C) |
The FSTest Salbutamol Rapid Test Kit is a qualitative, competitive colloidal gold lateral flow immunochromatographic assay engineered for rapid, on-site screening of salbutamol residues in animal muscle tissues.
Salbutamol is a potent beta-agonist compound prohibited or strictly regulated in livestock breeding globally due to food safety and consumer health concerns. With a high baseline sensitivity of 0.5 µg/kg (0.5 ppb) and clear visual interpretation within 8 to 10 minutes, this kit enables slaughterhouses, meat processors, livestock farms, and regulatory authorities to conduct efficient, dependable food safety monitoring without sophisticated chromatography equipment.
Ultra-Low Detection Limit (0.5 µg/kg LOD): Guarantees reliable screening at regulatory trace levels (0.5 ppb) across animal muscle tissues.
Fast Visual Readout in 8–10 Minutes: Delivers rapid, conclusive qualitative screening results, drastically shortening meat acceptance and release cycles.
Streamlined On-Site Workflow: Four simple steps (Extraction → Incubation → Sample Loading → Interpretation) eliminate the need for specialized laboratory personnel.
High Analytical Specificity: Demonstrated negative cross-reactivity with ractopamine at concentrations up to 0.5 mg/L.
Convenient Room-Temperature Storage: Stable at 4–30°C with an extended 18-month shelf life, significantly reducing cold-chain shipping and storage overhead.
| Product Name | Salbutamol Rapid Test Kit |
|---|---|
| Catalog Number | 26103 / DR32.4711.5 |
| Target Analyte Compound | Salbutamol (CAS: 18559-94-9 | C₁₃H₂₁NO₃ | 239.31 g/mol) |
| Detection Principle | Competitive Lateral Flow Immunochromatography (Colloidal Gold) |
| Pack Size | 10 tests / kit |
| Assay Run Time & Volume | 8–10 minutes (results invalid after 10 minutes); 100 µL test supernatant |
| Storage & Shelf Life | 18 months sealed at 4–30°C in a cool, dry place (Do not freeze) |
| Target Analyte Compound | CAS Registry Number | Molecular Formula | Limit of Detection (LOD) |
|---|---|---|---|
| Salbutamol (Beta-Agonist Residue) | 18559-94-9 | C₁₃H₂₁NO₃ | 0.5 µg/kg (0.5 ppb) |
Developed and validated specifically for food safety screening of beta-agonist residues in meat matrices:
STEP 1 Extraction Weigh 4.0 g homogenized tissue (fat/skin-free) into a 15 mL tube. Add 0.5 mL Extracting Solution, vortex 1 min, and incubate in an 80°C water bath for 10 min. | STEP 2 Centrifugation Centrifuge the heated tube at 4000 rpm for 3 minutes. Allow the resulting supernatant to cool down to room temperature for subsequent testing. | STEP 3 Microwell Incubation Ensure 20–40°C environment. Pipette 100 µL cooled supernatant into a microwell, mix up and down 3–5 times, and incubate undisturbed for 3 minutes. | STEP 4 Load & Read (8–10 Mins) Transfer all mixed solution from microwell into cassette sample well. Interpret visual lines strictly at 8–10 minutes. Results are invalid after 10 minutes. |
Negative (-): Both Test (T) line and Control (C) line appear clearly. T line color intensity is equal to or darker than C line → Salbutamol is absent or below 0.5 µg/kg.
Positive (+): Control (C) line appears distinctly, while Test (T) line is invisible or visibly weaker than C line → Salbutamol concentration ≥ 0.5 µg/kg.
Invalid: Control (C) line fails to appear, regardless of whether T line is visible. Re-test with a fresh cassette.
Adhere to strict operational parameters to ensure maximum screening accuracy:
Qualitative Screening Purpose: Engineered strictly for qualitative rapid screening; confirmatory quantitative testing should be performed via LC-MS/MS when required.
Cross-Reactivity Alert: This test card exhibits cross-reactivity with clenbuterol hydrochloride (samples containing clenbuterol hydrochloride may yield positive results); negative against 0.5 mg/L ractopamine.
Time-Sensitive Exposure: Once removed from the aluminum foil pouch, the test cassette must be used within 1 hour to prevent humidity degradation.
Strict Time Window: Read visual test results strictly between 8–10 minutes after loading cassette; results observed after 10 minutes are invalid.
Kit Components Included: 10 Test cassettes (individually sealed with desiccant), 10 Microwells (pre-coated reagent wells), Salbutamol Extracting Solution (7.5 mL) x1, 15 mL Centrifuge tubes x10, User Manual.
Required Equipment (Not Provided): Knives/scissors, tissue homogenizer, analytical balance, timers, 80°C water bath, vortex mixer, micropipettes (100 µL range) & tips, and centrifuge (≥ 4000 rpm).